Advertisement
Case Report| Volume 53, ISSUE 1, e29-e32, February 2018

Download started.

Ok

Transscleral fixation of a toric intraocular lens by a slipable suture technique

Published:August 29, 2017DOI:https://doi.org/10.1016/j.jcjo.2017.06.011
      In keratoplasty, astigmatic correction is crucial for achieving a favourable visual outcome. Several procedures have been introduced for astigmatic correction, including astigmatic keratotomy,
      • Hoffart L.
      • Proust H.
      • Matonti F.
      • Conrath J.
      • Ridings B.
      Correction of postkeratoplasty astigmatism by femtosecond laser compared with mechanized astigmatic keratotomy.
      wedge resection,
      • de la Paz M.F.
      • Sibila G.R.
      • Montenegro G.
      • et al.
      Wedge resection for high astigmatism after penetrating keratoplasty for keratoconus: refractive and histopathologic changes.
      and excimer laser refractive surgery.
      • Afshari N.A.
      • Schirra F.
      • Rapoza P.A.
      • et al.
      Laser in situ keratomileusis outcomes following radial keratotomy, astigmatic keratotomy, photorefractive keratectomy, and penetrating keratoplasty.
      Each of these procedures have limitations, including unpredictable refractive outcomes, technical difficulty, expenses, and potential complications.
      • Fares U.
      • Sarhan A.R.
      • Dua H.S.
      Management of postkeratoplasty astigmatism.
      The toric intraocular lens (TIOL) provides a surgical option for postkeratoplasty astigmatism. The development of a foldable TIOL that can be inserted through a small incision reduces the unpredictability of surgically induced astigmatism in sequential surgery after keratoplasty.
      • Fares U.
      • Sarhan A.R.
      • Dua H.S.
      Management of postkeratoplasty astigmatism.
      • Kersey J.P.
      • O’Donnell A.
      • Illingworth C.D.
      Cataract surgery with toric intraocular lenses can optimize uncorrected postoperative visual acuity in patients with marked corneal astigmatism.
      • Frohn A.
      • Dick H.B.
      • Thiel H.J.
      Implantation of a toric poly(methyl methacrylate) intraocular lens to correct high astigmatism.
      After combined cataract extraction and keratoplasty, the incompleteness of the can-opener capsulotomy or anteroposterior capsular adhesion precludes insertion of an IOL in the bag. Two cases of transscleral fixation of a TIOL for postkeratoplasty astigmatism have been reported,
      • McGrath L.A.
      • Lee G.A.
      Transscleral fixation of a supplementary toric intraocular lens to prevent rotation in a pseudophakic patient.
      • Borkenstein A.F.
      • Reuland A.
      • Limberger I.J.
      • Rabsilber T.M.
      • Auffarth G.U.
      Transscleral fixation of a toric intraocular lens to correct aphakic keratoplasty with high astigmatism.
      but in one, a supplementary sulcus TIOL was used,
      • McGrath L.A.
      • Lee G.A.
      Transscleral fixation of a supplementary toric intraocular lens to prevent rotation in a pseudophakic patient.
      and no solution for misalignment was suggested in the other.
      • Borkenstein A.F.
      • Reuland A.
      • Limberger I.J.
      • Rabsilber T.M.
      • Auffarth G.U.
      Transscleral fixation of a toric intraocular lens to correct aphakic keratoplasty with high astigmatism.
      Here, we describe a case that was successfully managed with a novel suture technique, which enabled readjustment of the haptic suture after scleral fixation, to resolve this complex surgical situation.
      A 66-year-old male presented with visual disturbance that had lasted several decades. He had experienced recurrent idiopathic keratitis. Visual acuity (by logMAR) was counting fingers OD and 1.2 OS. On slit-lamp examination, thick corneal opacities were observed bilaterally. The crystalline lenses showed significant opacities. A penetrating keratoplasty was performed with cataract extraction in the left eye. IOL implantation was not performed at that time because it is common practice to leave patients aphakic after keratoplasty and to perform a secondary IOL insertion to match the IOL power to the new keratometry values. Corneal sutures were removed after 1 year. Preoperative evaluation of the IOL implantation revealed a +3.00-diopter (D) sphere, +5.00-D cylinder, and a best corrected visual acuity of 0.2. The keratometric readings obtained via a manual keratometer (Ophthalmometer, OM-4; Topcon, Tokyo, Japan) were K1 = 48.50 D and K2 = 42.00 D, and the steep axis was 180°. Keratometric values of 49.50 and 42.37 D were obtained from a topographer (CT200; Dicon, San Diego, Calif.), and the steep axis was 2°. The axial curvature map showed a symmetric bowtie topography pattern of the left eye. Axial length was 25.57 mm, as determined by A-scan biometry (AXIS II PR Post Refractive Ultrasound; Quantel Medical, Toronto, Ont.). We calculated target power using the SRK-T formula as −0.25 D. An Acrysof IQ toric (Acrysof Toric Natural; Alcon Laboratories Inc, Fort Worth, Tex.) T9 +12.5-D IOL was selected for the secondary TIOL implantation. However, on slit-lamp examination, thick fibrous anteroposterior capsular adhesions were observed. Consequently, the patient could not undergo TIOL implantation in the bag. Instead, a transscleral IOL fixation was performed using a modified method.
      Before surgery, the structure of the TIOL was evaluated. The toric marks on the optic and the expected tying point of the haptic were identified by imaging software (LAS EZ 2.0; Leica Microsystems, Wetzlar, Germany). The angle between 2 imaginary axes, 1 formed by connecting the toric marks and the other by connecting the planned tying point, was calculated using ImageJ (National Institutes of Health, Bethesda, Md.). The nearer the tie to the proximal portion of the haptic, the greater the tension and torque on the suture. Therefore, we decided to place the tie at 2.0 mm from the end of the haptic, which would generate an angle of 40.7° from the toric axis (Fig. 1). Theoretically, two 3.0-mm-spaced polypropylene sutures could be tied on the haptic, enabling adjustment of the toric axis by 25.6° after tying.
      Fig. 1
      Fig. 1(A) Microscopic toric intraocular lens (IOL) measurements. The IOL (Acrysof Toric Natural T9; Alcon Laboratories Inc) was evaluated using a stereoscopic microscope (S6D; Leica Microsystems). Both the toric axis and the tying points were calculated via image analysis software (LAS EZ 2.0; Leica Microsystems). The toric axis is the imaginary axis passing through both toric marks in the IOL optic. The planned tying point (dark blue line) was placed 2.0 mm from the haptic end point. The planned tying point was at a 40.7° angle from the toric axis (red dashed line). (B) Theoretical principle rotation of the toric IOL. As the surgeon rotates the knot clockwise, the toric IOL is rotated counterclockwise. Using this manoeuvre, the surgeon can calibrate the toric axis to the steepest meridian.
      Immediately before the operation, the perpendicular axes were marked at the limbus. The steepest meridian (180°) and tying axis (40°) were marked. Two 3.0-mm scleral flaps were made at the 40° meridian. The capsule was removed using a 23G vitrector (Millenium Phaco Vitrectomy A/P CX3173; Bausch & Lomb, Bridgewater, N.J.). Two curved needles with a double-armed 10-0 polypropylene suture (Ethicon, Inc, Somerville, N.J.) were inserted 2.0 mm posterior to the limbus, separated by 3.0 mm, and passed from one scleral flap to the other with a docking technique using the 26-gauge needle. The 2 strands of the sutures were pulled out through the corneal incision with a Kuglen hook and cut between the strands. The new ends of the sutures on the same side were tied at one haptic, at 2.0 mm from the haptic end (as measured by caliper), using conventional square knots. After one haptic had been fixated with the sutures, the IOL was folded using Buratto forceps and was introduced into the anterior chamber via the same aforementioned clear corneal incision. The other haptic was externalized and tied in the same manner. By pulling on both ends of the sutures, the IOL was positioned in the eye and aligned toward the desired axis. Both ends of the sutures were tied to the scleral bed under the scleral flap, and the knots were locked. The suture loop at each haptic was mobilized and rotated with forceps, thus allowing the IOL to rotate as well. After positioning the IOL at the correct meridian, the viscoelastics and prolapsed vitreous fibers were removed. The overall surgical procedure is summarized in Video 1 (available online) and Figure 2.
      Fig. 2
      Fig. 2Intraoperative images (A–H) and corresponding schematic diagrams (I–P) of transscleral intraocular lens (IOL) fixation using a novel sliding knot technique. (A, I) The steepest meridian (arrows) was marked, and the haptic fixation meridian (arrow heads) was determined based on the steepest meridian. (B, J) A partial-thickness scleral flap was created using a microscalpel. (C, K) Each end of the double-armed 10-0 polypropylene suture was passed from the bed of the partial-thickness scleral flap 2.0 mm posterior to the limbus and 3.0 mm apart, using a 26G needle. (D, I) A 2.75-mm-wide clear corneal incision was made using a disposable scalpel. (E, M) The middle portion of the suture was extracted through a clear corneal incision. (F, N) The cut end of each suture was tied to the IOL haptics. (G, O) The outer portion of the suture was tied and rotated using microforceps. (H, P) The toric axis of the IOL was aligned to the steepest meridian.
      On postoperative day 1, uncorrected visual acuity was 0.5, and the IOL was well centred. Each toric mark was properly positioned at the planned meridian. The refraction was +2.00-D cylinder error at 180°. By 2 months, the keratometric readings were K1 = 49.00 D and K2 = 42.75 D, the steep axis was 8°, and refraction was +2.25-D cylinder error at 2°. Visual acuity had improved to 0.3. At 4 months, visual acuity remained at 0.3, and refraction was stable at +2.00-D cylinder at 4°. By 6 months, the TIOL was aligned to 180° as planned with the tying point stable at 40° on slit-lamp examination after pupil dilation. There were no suture-related complications.
      Recently, some studies have reported monofocal IOL implantation supplemented with a secondary TIOL in the sulcus.
      • Falzon K.
      • Stewart O.G.
      Correction of undesirable pseudophakic refractive error with the Sulcoflex intraocular lens.
      • Ferreira T.B.
      • Pinheiro J.
      Clinical results with a supplementary toric intraocular lens for the correction of astigmatism in pseudophakic patients.
      In the most challenging cases, however, such as a traumatic cataract with posterior capsule rupture or a radial tear caused by penetrating injuries, or patients presenting with capsular adhesions after simultaneous cataract extraction and keratoplasty, TIOL implantation is not a surgical option. In these cases, high astigmatism is one of the greatest obstacles to visual rehabilitation. In transscleral fixation, the TIOL should be fixed at the exact steepest meridian, but suturing of the haptic to the sclera can result in an improper axis. In our case, the capsular apparatus was distorted by adhesions. Therefore, a secondary TIOL implantation in the bag could not be performed, even though the patient had +5.00-D astigmatism.
      Previous reports have described surgical techniques
      • McGrath L.A.
      • Lee G.A.
      Transscleral fixation of a supplementary toric intraocular lens to prevent rotation in a pseudophakic patient.
      • Borkenstein A.F.
      • Reuland A.
      • Limberger I.J.
      • Rabsilber T.M.
      • Auffarth G.U.
      Transscleral fixation of a toric intraocular lens to correct aphakic keratoplasty with high astigmatism.
      • Emanuel M.E.
      • Randleman J.B.
      • Masket S.
      Scleral fixation of a one-piece toric intraocular lens.
      • Arjmand P.
      • Chan T.Y.
      • Ahmed I.I.
      Transscleral suture fixation following recurrent toric intraocular lens rotation.
      • Kelkar A.
      • Shah R.
      Tips and tricks of intrascleral fixation of single-piece and toric intraocular lens.
      that can be divided into 2 categories. The first is to reposition the TIOL in the capsular bag or ciliary sulcus because of malpositioning of the TIOL. Emanuel et al.
      • Emanuel M.E.
      • Randleman J.B.
      • Masket S.
      Scleral fixation of a one-piece toric intraocular lens.
      and Arjmand et al.
      • Arjmand P.
      • Chan T.Y.
      • Ahmed I.I.
      Transscleral suture fixation following recurrent toric intraocular lens rotation.
      reported in-the-bag suture fixation of the TIOL, and McGrath and Lee
      • McGrath L.A.
      • Lee G.A.
      Transscleral fixation of a supplementary toric intraocular lens to prevent rotation in a pseudophakic patient.
      reported scleral fixation of a malrotated add-on sulcus TIOL. However, these techniques are not applied as primary surgery for astigmatic correction because these techniques did not suggest a solution for the misalignment of TIOL even after the knots were tied and locked. The second category is primary scleral suture of the TIOL for correction of high astigmatism because the lens capsule lacks integrity. Borkenstein et al.
      • Borkenstein A.F.
      • Reuland A.
      • Limberger I.J.
      • Rabsilber T.M.
      • Auffarth G.U.
      Transscleral fixation of a toric intraocular lens to correct aphakic keratoplasty with high astigmatism.
      performed scleral sutures for the insertion of TIOL in a postkeratoplasty aphakic eye, and Kelkar and Shah
      • Kelkar A.
      • Shah R.
      Tips and tricks of intrascleral fixation of single-piece and toric intraocular lens.
      introduced scleral fixation using a sutureless technique. However, these techniques also did not suggest a strategy for the adjustment of the axis after fixation. Our technique is novel in that it allows calibration of the toric axis by means of sliding knots before completing the operation. This is useful, given the possibility of misalignment of the toric axis of the TIOL during its positioning and aligning toward the desired axis. We are not aware of previous descriptions of use of a sliding knot technique for adjusting the toric axis after fixation.
      In conclusion, using precise structural evaluation and a novel suturing technique, we here achieved a favourable outcome with transscleral fixation of a TIOL. By means of a sliding knot, the surgeon can adjust the toric axis to the planned steepest meridian after fixation. However, because a 1-piece IOL is not designed for haptic sutures as a prerequisite, the possibility of rotation, haptic or suture breaks, or slippage cannot be completely ruled out. Nevertheless, our study shows that transscleral fixation of a TIOL can be performed in highly astigmatic patients with an unstable or fibrotic capsular apparatus, using a sliding knot technique.
      This article includes online-only material. Video 1 can be found on the CJO web site at http://pubs.nrc-cnrc.gc.ca/cjo/cjo.html. It is linked to this article in the online contents of the xxx 20xx issue.

      Disclosure

      The authors have no proprietary or commercial interest in any materials discussed in this article.
      This work was supported by the National Research Foundation of Korea(NRF) Grant funded by the Korea Government(MSIP) (No. 2014R1A5A2010008).

      Appendix

      Supplementary Data

      References

        • Hoffart L.
        • Proust H.
        • Matonti F.
        • Conrath J.
        • Ridings B.
        Correction of postkeratoplasty astigmatism by femtosecond laser compared with mechanized astigmatic keratotomy.
        Am J Ophthalmol. 2009; 147 (e771): 779-787
        • de la Paz M.F.
        • Sibila G.R.
        • Montenegro G.
        • et al.
        Wedge resection for high astigmatism after penetrating keratoplasty for keratoconus: refractive and histopathologic changes.
        Cornea. 2010; 29: 595-600
        • Afshari N.A.
        • Schirra F.
        • Rapoza P.A.
        • et al.
        Laser in situ keratomileusis outcomes following radial keratotomy, astigmatic keratotomy, photorefractive keratectomy, and penetrating keratoplasty.
        J Cataract Refract Surg. 2005; 31: 2093-2100
        • Fares U.
        • Sarhan A.R.
        • Dua H.S.
        Management of postkeratoplasty astigmatism.
        J Cataract Refract Surg. 2012; 38: 2029-2039
        • Kersey J.P.
        • O’Donnell A.
        • Illingworth C.D.
        Cataract surgery with toric intraocular lenses can optimize uncorrected postoperative visual acuity in patients with marked corneal astigmatism.
        Cornea. 2007; 26: 133-135
        • Frohn A.
        • Dick H.B.
        • Thiel H.J.
        Implantation of a toric poly(methyl methacrylate) intraocular lens to correct high astigmatism.
        J Cataract Refract Surg. 1999; 25: 1675-1678
        • McGrath L.A.
        • Lee G.A.
        Transscleral fixation of a supplementary toric intraocular lens to prevent rotation in a pseudophakic patient.
        J Cataract Refract Surg. 2013; 39: 134-138
        • Borkenstein A.F.
        • Reuland A.
        • Limberger I.J.
        • Rabsilber T.M.
        • Auffarth G.U.
        Transscleral fixation of a toric intraocular lens to correct aphakic keratoplasty with high astigmatism.
        J Cataract Refract Surg. 2009; 35: 934-938
        • Falzon K.
        • Stewart O.G.
        Correction of undesirable pseudophakic refractive error with the Sulcoflex intraocular lens.
        J Refract Surg. 2012; 28: 614-619
        • Ferreira T.B.
        • Pinheiro J.
        Clinical results with a supplementary toric intraocular lens for the correction of astigmatism in pseudophakic patients.
        Eur J Ophthalmol. 2015; 25: 302-308
        • Emanuel M.E.
        • Randleman J.B.
        • Masket S.
        Scleral fixation of a one-piece toric intraocular lens.
        J Refract Surg. 2013; 29: 140-142
        • Arjmand P.
        • Chan T.Y.
        • Ahmed I.I.
        Transscleral suture fixation following recurrent toric intraocular lens rotation.
        J Cataract Refract Surg. 2015; 41: 912-917
        • Kelkar A.
        • Shah R.
        Tips and tricks of intrascleral fixation of single-piece and toric intraocular lens.
        J Clin Exp Ophthalmol. 2016; 7: 514